Abstract
Precise anesthesia delivery helps ensure amnesia, analgesia, and immobility. Conventionally, the end-tidal anesthetic concentration is maintained through manual adjustment of the fresh gas flow and anesthetic vaporizer output. Some anesthesia delivery systems can deliver and maintain clinician-selected end-tidal anesthetic agent (EtAA) concentration using a modified closed-loop system. We evaluated the performance of an End-tidal Control (EtC) system on the Aisys CS 2 anesthesia machine (GE HealthCare). We hypothesized EtC anesthetic delivery would be noninferior to manually controlled anesthetic delivery. The Multi-site Anesthesia randomized controlled STudy of End-tidal control compared to conventional Results (MASTER) Trial evaluated anesthetic delivery in 210 adult patients receiving inhaled anesthesia. Patients were randomized to either EtC or manual control (MC) anesthetic delivery. The primary objective was to determine whether, compared to conventional anesthesia practice, EtC achieves and maintains clinician-specified EtAA and end-tidal oxygen (Et o2 ) concentrations within defined noninferiority limits. Noninferiority was concluded if the lower limit of the 95% confidence interval (CI) of the difference between the percent duration within the acceptable range (5% of steady state or a margin of ~10% of each agent's minimum alveolar concentration) for EtC and MC was ≥ -5% for both EtAA and Et o2 . Secondary objectives included performance measures: response time: time required to attain 90% of the first desired EtAA, overshoot: amount the controller (or vaporizer delivery) exceeded the desired EtAA, and accuracy: average deviation from the desired EtAA. EtC achieved and sustained targeted EtAA and Et o2 concentrations within the noninferiority threshold. The EtAA was within 5% of the desired value 98% ± 2.05% of the time with EtC compared to 45.7% ± 31.7% of the time with MC (difference 52.3% [95% CI, 45.9%-58.6%], P < .0001). For Et o2 , EtC was within the noninferiority limit 86.3% ± 22.8% of the time compared with MC at 41% ± 33.3% ( P < .0001, difference 45.3% [95% CI, 36.1%-54.5%]). The median response time for achieving 90% of the initial EtAA desired value was 75 seconds with EtC and 158 seconds with MC ( P = .0013). EtC exhibited a median overshoot of 6.64% of the selected EtAA concentration, whereas MC often failed to reach the clinician's desired value. The difference in median percent deviation from desired EtAA value was 15.7% ([95% CI, 13.5%-19.0%], P < 0001). EtC achieves and maintains the EtAA and Et o2 concentration in a manner that is noninferior to manually controlled anesthesia delivery.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.