Abstract
Endotracheal tube (ETT) fire is a catastrophic disaster that may occur during laser surgery of the upper airway. Several means are available for protection of polyvinyl chloride (PVC) tube from fire, but they are not perfect in prevention of fires caused by laser beam. The PVC tube is hazardous for carbon dioxide (CO2) laser surgery if it is not well wrapped with metallized foil tape. We report a case that a PVC ETT wrapped with aluminum foil ignited during CO2 laser surgery of the larynx. In this report, we emphasize the shaft of the PVC tube must be completely wrapped with aluminum foil to prevent exposure of any surface if it is used in CO2 laser surgery of the upper aero digestive tract.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.