Abstract
An endoscopic pressure sensor has been evaluated in the measurement of oesophageal variceal pressure and its response to drug ingestion. The variceal pressure showed a highly significant correlation with the splenic pulp pressure (r = 0.97) in six patients with liver disease of diverse aetiology and with hepatic venous wedge pressure (r = 0.92) in eight alcoholic cirrhotic men. Intraduodenal infusion of isosorbide dinitrate in the cirrhotics produced no change in wedge pressure or endoscopic variceal pressure despite profound falls in arterial systolic pressure. Thus isosorbide dinitrate appears to be of no value in treating portal hypertension. This study establishes the endoscopic pressure sensor as a valuable tool in screening drugs in this condition.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.