Abstract

Endoscopic hand suturing (EHS) is a novel technique for closing a mucosal defect after endoscopic submucosal dissection (ESD). We investigated the technical feasibility of colorectal EHS using a modified flexible through-the-scope needle holder. This was a prospective multicenter study conducted at two referral centers between June 2022 and April 2023. This study included colorectal neoplasms 20-50 mm in size located in the sigmoid colon or rectum. A modified flexible through-the-scope needle holder, with an increased jaw width to facilitate needle grasping, was used for colorectal EHS. The primary end-points were sustained closure rate on second-look endoscopy (SLE) performed on postoperative days 3-4 and suturing time for colorectal EHS. Secondary end-points included complete closure rate and delayed adverse events. We enrolled 20 colorectal neoplasms in 20 patients, including four patients receiving antithrombotic agents. The tumor location was as follows: lower rectum (n = 8), upper rectum (n = 2), rectosigmoid colon (n = 4), and sigmoid colon (n = 6), and the median mucosal defect size was 37 mm (range, 21-65 mm). The complete closure rate was 90% (18/20 [95% confidence interval (CI) 68.3-98.8%]), and the median suturing time was 49 min (range, 23-92 min [95% CI 35-68 min]). Sustained closure rate on SLE was 85% (17/20 [95% CI 62.1-96.8%]). No delayed adverse events were observed. EHS demonstrated a high sustained closure rate. Given the long suturing time and technical difficulty, EHS should be reserved for cases with a high risk of delayed adverse events.