Abstract
A flexible needle-catheter tissue-anchoring device was developed to accomplish full-thickness tissue apposition of the GI wall. The aim of this study was to identify the performance of this device for repair of large iatrogenic gastric perforations in a porcine model. Six pigs. Short-term survival animal study. Pigs were studied while they were under general anesthesia. Device performance in differing gastric locations and wall thicknesses was assessed by 2 perforations more than 2 cm in size created for each pig along the greater curvature and the anterior wall. Each perforation was closed by parallel placement of tissue anchor sets sequentially along the length of the perforation. One week follow-up endoscopy and necropsy were performed. Twelve perforations were closed with the 48 tissue anchor sets. All animals survived for 1 week without clinical complications. Follow-up endoscopy and necropsy revealed that all tissue anchors remained with firmly held sutures and sealed perforations. Full-thickness closure with a new tissue-anchoring device simply and successfully repaired large iatrogenic gastric perforations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.