Abstract

After bilateral implantation of an angle-supported anterior chamber intraocular lens (AC IOL) in an allergic patient, the haptic tips were exposed through the limbus in the left eye. Endophthalmitis developed in the eye and was treated by pars plana vitrectomy, IOL extraction, and intravitreal antibiotic injection. Haptic repositioning was performed in the right eye. In this report, we assess the possibility that exposure of the angle-supported AC IOL haptics caused the endophthalmitis. We hypothesized that vigorous rubbing of the eye of the atopic patient might cause limbal erosion that could lead to exposure of the angle-supported AC IOL haptic through the limbus.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.