Abstract

The authors attempted ureteral occlusion by means of heat application in nine ureters (24 sites) of New Zealand White rabbits with the electromagnetic field-focusing (EFF) device. The EFF device generates heat at the tip of a grounded probe by focusing eddy currents that have been induced within the tissues by an external radio-frequency field. The power settings were varied from 30 to 150 W. Heat was applied at multiple sites in each ureter. Immediate functional occlusion was seen in all nine ureters. Long-term complete occlusion was seen in six ureters at power settings ranging from 40 to 150 W, while long-term partial occlusion was seen in two ureters at 30-50 W. All sites at 30 W resulted in partial occlusions. Perforation of the ureter resulted in urinoma formation in one ureter at a site that was treated with 150 W. The EFF device can be used to endoluminally occlude the ureter by causing a fibrotic reaction to thermal injury. The effective power range for this application appears to be 40-100 W.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call