Abstract

Specially developed high-resolution real-time ultrasound transducers (12.5 and 20 MHz) on the tip of endoluminal catheters were inserted into the endometrial canal to evaluate the usefulness of this approach. Uterine abnormalities, most confirmed by biopsy, surgery, or both, were detected in 12 patients, including submucosal myomas, nabothian cysts, endometrial polyps, synechiae, and endometrial and cervical carcinoma. In one case the catheter was directed under hysteroscopic guidance into a fallopian tube, demonstrating its potential usefulness in this region. In 4 of the 12 cases in which a hysterectomy was performed, an in vitro ultrasound examination of the organ was performed, which confirmed the initial in vivo ultrasound impressions. Anatomic cross-sectional slices of the uterus resulted in excellent correlation with the ultrasound findings. These preliminary results suggest that this new sonographic procedure will become an important diagnostic tool, supplementing abdominal and endovaginal ultrasound approaches.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.