Abstract
We describe our experience with endoluminal repair of abdominal aortic aneurysms using the stent-graft device. Twenty-four patients underwent 25 procedures in the 27-month period ending December 31, 1992. Twenty-one of the patients were considered high-risk candidates for conventional surgical repair. The endoluminal stented grafts were aortoaortic in 16 procedures and unilateral aortoiliac in eight. One patient underwent a second procedure consisting of an ilioiliac graft to repair a separate common iliac artery aneurysm. Technical problems were primarily related to retrograde transluminal access across the iliac arteries, tortuous aneurysms, and misjudgments as to measurement of length. One patient died and another required secondary deployment of a distal stent at 4 months; subsequent aneurysm expansion mandated surgical replacement at 18 months. It is clear that this device and methodology will have to undergo further refinement before the technique is acceptable for wider clinical application. Current experience, however, is encouraging. Aneurysm exclusion with an endoluminal prosthesis is likely to become an important therapeutic alternative over the next several years.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.