Abstract

ObjectiveIn endovascular aortic aneurysm repair (EVAR), proximal type 1A endoleaks can occur as a result of hostile neck anatomy or over- or undersizing of the endograft. As the current standard is based on the diameter or average of the short and long axes in a central lumen reconstruction image, it can falter in irregularly shaped aortic necks. An alternative method is circumference-based, therefore minimizing the measurement error. In this study we aimed to assess the degree of discrepancy between both methods and the association of this discrepancy with the occurrence of endoleak type 1A.MethodsAll patients with early (<30 days post-operative) endoleak type 1A after elective EVAR at our center between 2004 and 2016 were identified for a retrospective case-control study. Control patients were matched based on hostile neck anatomy, such as calcification, thrombus, reverse taper, and β-angulation. The aortic neck diameter was measured using the traditional, diameter-based method as well as an alternative method, based on the circumference of the aortic neck.ResultsIn 482 EVAR patients, 18 early endoleak type 1A cases were found (3.9%). After exclusion, 12 cases remained and 48 matching controls were found. No significant differences were found between the two measuring methods at any level below the renal arteries. The inter-observer variability was significant for the D(mean) (0.4 ± 1.69 mm, P = .02) and was larger than the D(circ) method (-0.1 ± 1.03 mm, P = .35). In only four out of 12 cases the endograft size was 10–20% larger than the D(mean) and D(circ) measurements. The differences between the diameter of the D(mean) and D(circ) and the chosen endograft were smaller for the case group (-8 ± 25.6% and -7 ± 24%) than for the control group. (-12.4 ± 12.4% and -11 ± 10.7%).ConclusionThe difference between the D(mean) and D(circ) methods for aortic neck measurement was not large enough to play a significant role in the incidence of endoleak type 1A. Inadequate oversizing and considerable β-angulation of the aortic neck may have been the cause of endoleak type 1A in this population. Robust and well-investigated sizing methods are paramount for accurate endograft sizing and prevention of endoleak type 1A. Therefore the lack of studies in this field and a sizeable inter-observer variability do not justify the widespread reliance on the traditional diameter-based methods for endograft sizing.

Highlights

  • The introduction of endovascular techniques for the treatment of abdominal aortic aneurysms (AAA) has lead to a decrease in perioperative mortality

  • In endovascular aortic aneurysm repair (EVAR), proximal type 1A endoleaks can occur as a result of hostile neck anatomy or over- or undersizing of the endograft

  • The differences between the diameter of the D(mean) and D(circ) and the chosen endograft were smaller for the case group (-8 ± 25.6% and -7 ± 24%) than for the control group. (-12.4 ± 12.4% and -11 ± 10.7%)

Read more

Summary

Introduction

The introduction of endovascular techniques for the treatment of abdominal aortic aneurysms (AAA) has lead to a decrease in perioperative mortality. In the case of endoleak type 1A, the endograft does not completely seal the proximal aneurysm neck and arterial flow is present between the wall of the aortic neck and the graft material. This flow may lead to further growth and eventually rupture of the aneurysm. Hostile neck characteristics have been associated with the occurrence of endoleak type 1A These characteristics include a short proximal aneurysm neck length, reverse tapering of the neck, mural calcification or thrombus, and severe neck angulation [5]

Methods
Results
Discussion
Conclusion