Abstract

The Endocuff is a new colonoscopy accessory that has been designed to improve both the adenoma detection rate and endoscope tip control. A total of 50 Endocuff-assisted colonoscopies were analyzed retrospectively with regard to safety, procedural success, and complications. The cecal intubation rate was 98 %, and the mean intubation time was 6.0 minutes (95 % confidence interval 5.3 - 6.6 minutes). The ileal intubation rate was 76 %. In 30 % of patients, the Endocuff caused small, superficial, "scratch-like" mucosal lesions. In all other patients, no Endocuff-associated complications were observed. A total of 36 adenomas were detected in 50 patients. The adenoma detection rate was 34 %. Endocuff-assisted colonoscopy showed good procedural success rates in terms of cecal intubation rate and time, and a promising adenoma detection rate. Endocuff seems to improve endoscope tip control, especially during polypectomy. Endocuff may be a useful device for colorectal adenoma screening, and should be investigated in larger trials.