Abstract

The study is intended to perform an end-to-end test of the entire intraoperative process using cadaver heads. A simulation of tumor removal was performed, followed by irradiation of the bed and measurement of absorbed doses with radiochromic films. Low-energy X-ray intraoperative radiotherapy (IORT) was used for irradiation. A computed tomography study was performed at each site and the absorbed doses calculated by the treatment planning system, as well as absorbed doses with radiochromic films, were studied. The absorbed doses in the organs at risk (OAR) were evaluated in each case, obtaining maximum doses within the tolerance limits. The absorbed doses in the target were verified and the deviations were <1%. These tests demonstrated that this comprehensive procedure is a reproducible quality assurance tool which allows continuous assessment of the dosimetric and geometric accuracy of clinical brain IORT treatments. Furthermore, the absorbed doses measured in both target and OAR are optimal for these treatments.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.