End tidal carbon dioxide measurement with an electro-acoustic sensor is demonstrated. The sensor consists of an acoustic resonator coupled to a low cost electro-acoustic element. By simultaneous measurements with a reference sensor, the new device was tested on subjects performing exercise, hypo- and hyperventilation whereby the CO<inf>2</inf>concentration ranged from 2.1 to 7.0 kPa. The output from the experimental device correlated well with the reference CO<inf>2</inf>readings with a correlation coefficient of 0.976. Response time for expiration less than 0.8 seconds was noted. The new device could be useful in situations where selectivity to other gases is not important.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.