Abstract
End-on observation of a horizontal water-cooled low-flow plasma has been compared to side-on observation of a conventional high flow plasma. The analytical performance is similar to the conventional plasma. The detection limits are close to the detection limits measured with a Perkin-Elmer Plasma II conventional ICP. The linearity of calibration curves is also similar to the conventional ICP. However, a remarkable reduction of interferences has been observed for end-on observation of the low-flow plasma.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.