Abstract

Three encoding strategies for the measurement of flow velocities in arbitrary directions with phase-contrast magnetic resonance imaging are presented; their noise and dynamic range performance are compared by means of theoretical analysis and computer simulation. A six-point measurement strategy is shown to be quite inefficient in terms of velocity variance per unit time. A simple four-point method exhibits equal dynamic range; its noise depends on flow direction but on average is equal to that of the six-point method. An alternate, balanced four-point method has noise that is direction independent and has, depending on implementation, possibly lower noise levels. Either four-point method is more efficient and is preferred over the six-point approach.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.