Abstract
The encapsulation of high purity germanium detectors is very desirable in order to increase their versatility and reliability. However, rapid and extensive degradation is seen for all detectors made from detector grade crystals which are encapsulated in a simple vacuum. Extensive studies have shown that the cause of this degradation is hydrogen adsorption on the detector surface. There it causes the formation of a strong p+ inversion layer which in turn shunts the detector junction. Reliable encapsulation is shown to be achievable by the use of hydrogen-free germanium crystals.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.