Abstract

Capillary array electrophoresis (CAE) is a promising technique for multiple enantiomeric separations. Carboxytetramethylrhodamine succinimidyl ester (TAMRA SE), a rhodamine-core fluorescent probe, has rarely been applied as an original precolumn derivatization reagent for chiral amino acid (AA) analysis so far. For these purposes, high-throughput enantiomeric separations of 12 TAMRA SE-AAs by a home-made 532nm CAE-LIF scanner are presented. The effect of cyclodextrins (CDs) and a variety of organic modifiers was quickly investigated. Baseline separations were achieved in 100mM Tris–borate buffer (pH 10.0) containing 2mM β-CD and 10mM hexamethylenediamine (HDA). Multiple determination of the enantiomeric excess (ee) in non-racemic mixtures of alanine is successfully presented.