Abstract

Cloprostenol (CLO) is an important chiral drug widely used in veterinary practice. In high-performance liquid chromatography (HPLC), the acetonitrile/water system is more effective in the chiral resolution of CLO. The change in the ratio of acetonitrile and water has a great influence on the chiral selectivity of CLO. The enantioselective adsorption and nonselective adsorption of CLO on the polysaccharide stationary phase with three groups of mobile phases (Acetonitrile / 0.1 TFA; 95 Acetonitrile/5 H2O/0.1 TFA; 45 Acetonitrile/55 H2O/0.1 TFA) were determined and fitted with the subtraction fitting method (SFM). Adding a small amount of water to the mobile phase mainly reduced the enantioselective equilibrium adsorption constants Ks of (-)-CLO and thus changed the selectivity. Among them, the true separation factor (αtrue) of the 95 Acetonitrile/5 H2O mobile phase was as high as 92.86. Chiral preparation was performed on the basis of this mobile phase. The preparation performance in kkd (kilograms per kilogram of stationary phase purified compound per day) was 0.25 kg racemate/kg CSP/day.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call