Abstract

A simple CE method has been developed for the separation and determination of thyroxine (T4) enantiomers in pharmaceutical formulations. The method was based on ligand-exchange mechanism using a Cu(II)/L-proline complex as chiral selector. The effects of different parameters affecting separation such as chiral selector concentration, organic additive, buffer pH and temperature were investigated. A baseline separation of the two enantiomers was obtained at a Cu(II)/L-proline ratio of 1:8 in a borate buffer (15 mmol/L, pH 9.6) containing 10% v/v acetonitrile. Under the optimized conditions, precision linearity range and detection limits of the developed enantioselective CE method were evaluated and compared using two different detection systems: conventional UV detection at 226 nm and iodine (127I)specific detection ("chiral speciation") with ICP-MS. Both methodologies show adequate analytical performance characteristics with detection limits around 0.30 microg/mL for each enantiomer of T4. Finally, a levothroid pharmaceutical formulation sample was successfully analyzed using both developed methods CE-UV and CE-ICP-MS.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.