Abstract

A new stereospecific LC method for the separation and quantification of moxifloxacin and its (R,R)-enantiomer in bulk drug was developed and validated by ligand-exchange liquid chromatography on a reversed phase column using aqueous mobile phase containing the chiral reagent l-isoleucine-Cu(II). The UV detector was operated at 293 nm. The flow rate of mobile phase was set at 0.9 mL min−1. The achiral ODS column offers good separation of the two enantiomers in less than 20 min. The test concentration was 1,000 μg mL−1 in the mobile phase. This method was capable of detecting the (R,R)-enantiomer of moxifloxacin up to 0.1 μg mL−1 for a 20 μL injection volume. The drug was subjected to stress conditions of hydrolysis, oxidation, photolysis and thermal degradation. There was no interference of degradants with the (R,R)-enantiomer in the developed method. The developed chiral RP-LC method was validated with respect to linearity, accuracy, precision and robustness. The percentage recovery for the (R,R)-enantiomer in bulk drug samples ranged from 98.1 to 104.4%. The test solution was found to be stable in the mobile phase for 48 h after preparation.