Abstract
Capillary electrophoresis (CE) coupled to tandem mass spectrometry was applied to the chiral separation of baclofen using sulfobutylether-β-cyclodextrin chiral selector in partial filling counter current mode. On-line UV detection was simultaneously used. Method optimization was performed by studying the effect of cyclodextrin and BGE concentration as well as sheath liquid composition on analyte migration time and enantiomeric resolution. The cyclodextrin showed stereoselective complexation towards baclofen enantiomers, allowing chiral resolution at low concentration. The CE capillary protrusion from the ESI needle relevantly affected the chiral resolution and the analyte migration time. Complete enantiomeric separation was obtained by using 0.25 M formic acid BGE containing 1.75 mM of chiral selector and water/methanol (30:70, v/v) 3% formic acid as sheath liquid. The method exhibited a LOD of 0.1 μg/mL (racemic concentration) in MS 3 product ion scan mode of detection and was applied to the analysis of racemic baclofen in pharmaceutical formulations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.