Abstract

The separation and quantitative determination of atenolol isomers by chiral high-performance liquid chromatography (HPLC) are described. Atenolol isomers were separated using a Chiralcel OD® column (250 × 4.6 mm,10 μm); the mobile phase was hexane-ethanol-diethylamine (75:25:0.1 v/v/v); ultraviolet detection was at 276 nm; and flow rate was 0.7 ml/min. The coefficient of variation and average recovery of (R)-isomer were 0.60% and 100.37%, respectively, for sample A and 0.69% and 100.63%, respectively, for sample B. The coefficient of variation and average recovery of (S)-isomer were 0.59% and 100.33%, respectively, for sample A and 0.63% and 99.78%, respectively, for sample B.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.