Enantiomer labelling, a method for the quantitative analysis of opticaly active natural compounds by gas chromatography, involves theuse of the unnatural enantiomer as an internal standard. WithChirasil-Val, a chiral stationary phase that is thermally stable up to up to 240°, the enantiomers of amino acid and a variety of other compounds can be separated and quantitated. Incomplete recovery from the sample, incomplete derivatization, hydrolysis and thermal decomposition of the derivative and shifting response factors can be compensated for by adding the unnatural enantiomer. The accuracy of amino acid analysis by enantiomer labelling is equal or superior to that of hitherto known methods. The procedure affords a complete analysis of peptides with respect to both amino acid composition and the optical purity of each amino acid.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.