Abstract

Establishing and maintaining the correct formulation composition is essential for ensuring the stability of biopharmaceutical drug products. A barrier to the routine assessment of excipient concentration is the lack of convenient and robust methods for the direct analysis of solutions containing high protein concentrations. To address this need an HPLC method was developed utilizing a wide-pore C18 guard column to trap proteins in-line with a hydrophilic interaction liquid chromatographic column to separate excipients. This method allows for a simple and direct analysis of excipients such as amino acids, carboxylic acids, simple carbohydrates, and inorganic ions across multiple drug formulations and process streams containing different therapeutic antibodies. The method was successfully validated for specificity, precision, accuracy, linearity, and robustness.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.