Abstract
The increase histopathological evaluation of prostatectomy specimens rises the workload on pathologists. Automated histopathology systems, preferably directly on unstained specimens, would accelerate the pathology workflow. In this study, we investigate the potential of quantitative analysis of optical coherence tomography (OCT) to separate benign from malignant prostate tissue automatically. Twenty fixated prostates were cut, from which 54 slices were scanned by OCT. Quantitative OCT metrics (attenuation coefficient, residue, goodness-of-fit) were compared for different tissue types, annotated on the histology slides. To avoid misclassification, the poor-quality slides, and edges of annotations were excluded. Accurate registration of OCT data with histology was achieved in 31 slices. After removing outliers, 56% of the OCT data was compared with histopathology. The quantitative data could not separate malignant from benign tissue. Logistic regression resulted in malignant detection with a sensitivity of 0.80 and a specificity of 0.34. Quantitative OCT analysis should be improved before clinical use.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.