Abstract
A reflectance spectroscopic device that utilizes a single fiber for both light delivery and collection has advantages over classical multi-fiber probes. This study presents a novel empirical relationship between the single fiber path length and the combined effect of both the absorption coefficient, mua (range: 0.1-6 mm-1), and the reduced scattering coefficient, micro's (range: 0.3 - 10 mm-1), for different anisotropy values (0.75 and 0.92), and is applicable to probes containing a wide range of fiber diameters (range: 200-2000 microm). The results indicate that the model is capable of accurately predicting the single fiber path length over a wide range (r = 0.995; range: 180-3940 microm) and predictions do not show bias as a function of either microa or micro's .
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.