Abstract
This concept article reports a practical solution to improve the linearity of effluent pH response as observed in pH gradient cation exchange chromatography (CEX). When performing pH gradient CEX, it is not easy to develop buffer systems that will universally provide pH response proportional with the mobile phase (buffer) composition. It is an especially challenging pursuit when exploring MS compatible buffers (e.g. ammonium-acetate, ammonium-carbonate). In addition to „non-proportional” behavior from the mobile phase composition, the chromatographic column itself will sometimes impose an unpredictable impact on the effluent pH. Here, we propose a simple approach based on the on-line measurement of effluent pH response, conversion of pH to mobile phase volume fraction (φ) and then generation of the inverse response function in the time domain. In the end, when setting the inverse function as the gradient program instead of a linear gradient, an improved – ideally linear - pH response can be produced. A simple Excel tool was developed to assist analysts with this correction procedure, and it has been made available by download for public use.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.