Abstract

BackgroundThis paper proposes a new approach for neural control of hand prostheses, grounded on pattern recognition applied to the envelope of neural signals (eENG). New methodThe ENG envelope was computed by taking into account the amplitude and the occurrence of the spike in the neural recording. A pattern recognition algorithm applied on muscular signals was defined as a reference and a comparative analysis with traditionally adopted Spike Sorting Algorithms (SSA) for neural signals has been carried out. Method validation was divided in two parts: firstly, neural signals recorded from one amputee subject through intraneural electrodes were offline analyzed to discriminate between the two performed gestures; secondly, algorithm performance decay with the increase of the number of classes was studied through synthetic data. ResultsAn accuracy of 98.26% with real data was reached with the pattern recognition applied to eENG. SSA reached an accuracy of 70%. Increasing the number of classes worsens the accuracy of this algorithm. Additionally, computational time for the pattern recognition applied to eENG is very low (32.6 μs for each sample in the data window analyzed). Comparison with existing methodThe eENG was proved to be more reliable in decoding the user intention than the SSA algorithm and it is computationally efficient. ConclusionsIt was demonstrated that it is possible to apply the well-known techniques of EMG pattern recognition to a conveniently processed neural signal and can pave the way to the application of neural gesture decoding in upper limb prosthetics.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call