Abstract

Infant incubator is one of MEEs (Medical electrical equipment) that is largely used to preserve preterm and sick babies from postnatal stressors. However, their motors produce high electromagnetic fields (EMFs) which could affect the incubator as well as its occupants. In this paper, it has been done the EMF measurement to study and to quantify the level of EMF by mapping the magnetic field distribution inside the compartment of infant incubator. Measurement method was conducted on three layers from the mattress by magnetic near field probe and Spectrum Analyzers. The result showed that by p<0.005, the EMF is not significantly different from layer 1, layer 2, and layer 3. The averages values of the EMF are 0.53 <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$\mu \mathrm{T}, 0.98\mu \mathrm{T}$</tex> , and 0.97 <tex xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">$\mu \mathrm{T}$</tex> , at first, second and third layer on every stage of heater level. This EMF is far below the standard level for magnetic field ERLs for exposure of limbs, and head and torso at 0 Hz to 5MHz by the IEEE and ICNRP.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.