Abstract

Human finger- and toenails have been tested with an X-band EPR technique for different conditions of nail storage. The main radiation-induced signal at g = 2.005 demonstrated good stability if the samples were stored in a vacuum at room temperature after nail harvesting and irradiation. On the basis of this phenomenon, a new protocol is proposed to use the nails as possible emergency EPR dosimeters. The dosimetry protocol was tested on laboratory-exposed samples and demonstrated the ability to recover doses in the region 0–10 Gy with an estimated uncertainty of approximately 0.3–0.4 Gy for doses in the range <2 Gy, increasing to 0.6–0.7 Gy for doses in the range 5–10 Gy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.