Abstract
A 44-year-old woman with a history of transient ischemic attack underwent closure of atrial septal defect with a 26 mm Amplatzer device. The device was released without residual shunt or impingement on intracardiac structures. Within seconds, the transesophageal echocardiography showed the initial dislodgement of the device from the atrial septum and its consequent slipping back into the right atrium close to the tricuspid valve. Soon after the device disappeared from the right atrium and it could be founded into the right ventricle under the tricuspid valve. The patient was transferred in the operating room for an emergency operation. The device could not be found in the right ventricle because its downstream migration. The Amplatzer septal occluder was identified by palpation into the pulmonary artery trunk: it was retrieved from the right ventricle through the pulmonary valve and the atrial septal defect was closed by running suture.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.