Abstract
Measuring time-resolved spectra is crucial in inertial confinement fusion and radiation source development experiments. An elliptically bent crystal spectrometer is designed to measure X-rays in the range of 2.5-11.0 keV, which was achieved using four different lattice spacings of 0.8512, 0.6687, 0.4246, and 0.2749 nm with spectral resolution E/δE of ∼500. The X-rays emitted from a source at one focus of the ellipse undergo Bragg reflection off a crystal and pass through the second focus of the ellipse to a streak camera slit with 18-mm length and 80-μm width to generate a time-resolved spectrum. An alignment method for the time-resolved spectrometer was developed with the straight line connecting the centers of the two small holes on the fabricated substrate being the axis of the ellipse, thus allowing the spacing between the source and the elliptical crystal to be tuned to couple with the streak camera. The time-resolved spectrometer's performance was experimentally tested at the Shenguang II laser facility. The results indicate that its performance is close to that predicted theoretically.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.