Abstract
Solid-phase microextraction coupled to high-performance liquid chromatography (SPME–HPLC) with fluorescence detection was employed to determine bisphenol A (BPA) in milk samples. The potential influence of the milk matrix on the determination of BPA by SPME–HPLC were investigated. Optimal conditions to eliminate any matrix effects were as follows: milk samples were deproteinized with trichloroacetic acid, diluted 20-fold with BPA-free Ultrapure water, dissolved in methanol, the precipitated protein was filtered out, rinsed with methanol and evaporated to remove the methanol. Then, a 40.0-ml solution was used for SPME extraction and HPLC analysis. Satisfactory recoveries (milk: 93.1–101%; soybean milk: 93.9–102%) were achieved. The proposed method was successfully applied to real samples, BPA being detected within the range 1.6–2.6 ng ml−1 in four brands of commercial milk but not in soybean milk.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.