Abstract

Contaminants on the boundary surfaces lead to electric potential inhomogeneities. The resulting poorly defined electric force leads to ambiguities in precision Casimir force measurements. We experimentally demonstrate that with UV light followed by Ar ion beam radiation the electrostatic effects from these patch potentials can be eliminated. The cleaning procedure discussed reduces the sphere-plate potential difference to near zero. In addition, the UV radiation results in vacuum chamber pressures being reduced by a factor of 10 from the removal of volatile molecular species. This leads to very stable and near zero sphere-plate potential differences. The reported combination UV and Ar ion in situ cleaning procedure will find wide application.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.