Abstract

In this prospective randomized study, we investigate metal ion concentrations and clinical outcome 2 years after metal-on-metal (28 patients) or metal-on-polyethylene (26 patients) hip arthroplasty with 28-mm modular heads. Metal ion concentrations in patient serum were analyzed by high-resolution plasma mass spectrometry. The clinical outcome was almost identical in both groups with respect to the Harris hip score and the Medical Outcome Study Short Form–36; and on plain radiography, no signs of loosening occurred in any group. In the metal-on-metal group, concentrations of all investigated ions increased significantly when compared with preoperative values: cobalt, 15.3-fold (95% confidence interval [CI], 9.4-21.2); chromium, 5.2-fold (CI, 3.5-7.0); nickel, 2.1-fold (CI, 1.2-3.0); and manganese, 1.6-fold (CI, 1.3-2.0). In the metal-on-polyethylene group, a smaller but significant increase in the concentrations of cobalt and nickel occurred, whereas concentrations of the other ions did not change significantly.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.