Abstract

Using AC spark, AC arc and DC arc excitations in optical emission spectroscopic systems, the Al metal used in reactors can be characterised for its minor or trace element composition. The best precision of ± 6% is obtained with an AC spark in which the rod sample is taken as a self electrode and elements Cu, Fe, Mn, Si and Ti are determined in the concentration range 0·01–1%. The oxide powder sample with DC arc excitation provides best minimum detection limits of 10 ppm in general and 21 trace elements are determined by it. The AC arc method also uses oxide powder standards prepared synthetically and determines B and Mg in addition to the elements determined in AC spark excitation with a precision of ±9%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.