Abstract
Trace elemental analysis using microwave-induced plasma (MP) generated by nitrogen gas was employed as an atomization and excitation source for emission spectrometric (microwave plasma atomic emission spectroscopy, MP-AES) analysis for crude oil samples. Nitrogen gas produced from air through a gas generator, and in combination with an external gas control module (EGCM) used to introduce air into the plasma, leads to a stable robust microwave-induced plasma, enough to perform the analysis of crude oil samples with different API values, sulfur (0.5–5 wt %) and nitrogen (500 −2500 mg/kg) by direct dilution in a o-xylene diluent. Excellent detection limits and spike recoveries at low and high concentration levels were determined for Ni, V, Fe, Ca, and Na in crude oil matrix. The recoveries obtained from the analysis of the three quality control (QC) test materials were within ±10% of the actual/certified values. Comparable results for Ni and V as well as the V/(V + Ni) ratio in crude oils between nitrogen pl...
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.