Abstract

Laser ablation-inductively coupled plasma-mass spectrometry (LA-ICP-MS) was used as a complementary technique to X-ray fluorescence (XRF), for multi-element analysis of geological samples fused with lithium-tetraborate Li2B4O7). Different calibration strategies using external non-matrix matched reference materials were investigated. Various internal standards were tested, including the use of Li from the flux, and the use of the naturally occurring internal standards, Si or Ca. The use of a naturally occurring standard is not as efficient as this required a prior analysis of the samples using XRF. The obtained values for the analysis of geological reference materials were compared with consensus literature values, and satisfactory agreement was found. Laser pits, which were formed, had a diameter of 80 μm and 3–5 replicates on each fused disc were measured. The reproducibility of the method was better than 10% for concentrations above 1 μg/g and better than 15% for lower concentrations. The use of Li as the internal standard offers the possibility of multi-element determinations in geological samples, which have an unknown composition when the laser ablation analysis is carried out. However, using the calculated stoichiometric composition of the lithium-tetraborate for the calculation of the Li concentration leads to a constant deviation from the recommended values. Therefore, it was necessary to determine the Li concentration within each sample batch using at least one lithium tetraborate fused geological reference material. This resulting Li concentration in the beads was then used for all subsequent samples in a run. Limits of detection, reproducibility, deviation from reference values indicate the potential of LA-ICP-MS for such bulk analysis without matrix matched calibration standards.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call