Abstract

A simple and inexpensive electrothermal vaporization device consisting of a double-layer tungsten coil, of the type normally manufactured for halogen lamps, was used for the simultaneous determination of Al, Ca, Cr, Cu, Fe, Mg, Mn, Ni and Ti in aqueous suspensions of silicon carbide powder by inductively coupled plasma atomic emission spectrometry (ICP-AES). Possible interferences were investigated and background correction are discussed. Excluding Al, the limits of detection achievable were at the sub-µg g–1 and µg g–1 level. The accuracy was checked by comparison of the results with those obtained by instrumental neutron activation analysis (INAA), slurry sampling electrothermal atomic absorption spectrometry and ICP-AES involving decomposition of the sample. The precision expressed as relative standard deviation was between 3.3%(for 210 µg g–1 of Ti) and 13.5%(for 8.9 µg g–1 of Ni).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.