A solid sampling atomic absorption method for the determination of Cu, Fe, Mn, Na and Zn in high-purity tantalum powders has been developed. It is based on a direct automatic introduction of the sample into the graphite tube using the platform technique. Standard hollow-cathode lamps in combination with deuterium background correction are used as radiation sources. This and some additional modifications of the Grün spectrometer hard- and software lead to a significant improvement of its performance. With the maximum applicable sample mass of 40 mg limits of detection of 1 ng g −1, 27 ng g −1, 1.5 ng g −1, 0.7 ng g −1 and 0.1 ng g −1, respectively, are achieved for Cu, Fe, Mn, Na and Zn. Aqueous standard solutions are used for calibration. Comparison of the results obtained by solid sampling atomic absorption spectrometry with those of the liquid sampling technique show satisfactory agreement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.