Abstract

Electrosurgical currents can be delivered to tissue in monopolar or bipolar and monoterminal or biterminal modes, with the primary difference between these modes being their safety profiles. A monopolar electrosurgical circuit includes an active electrode and a dispersive (return) electrode, while there are 2 active electrodes in bipolar mode. In monoterminal mode, there is an active electrode, but there is no dispersive electrode connected to the patient's body and instead the earth acts as the return electrode. Biterminal mode uses a dispersive electrode connected to the patient's body, has a higher maximum power, and can be safer than monoterminal mode in certain situations. Electrosurgical units have different technologies for controlling the output power and for providing safety. A thorough understanding of these technologies helps with a better selection of the appropriate surgical generator and modes.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.