Abstract
The translation of impedance (R), current (I), and voltage (V) into tissue effects and the understanding of the settings of electrosurgical units is not obvious if judged by the many questions during live surgery. Below 200 V, the current heats the tissue until the steam of boiling stops the current. Thus, slower heating, because of less energy or a larger contact area, results in deeper coagulation. Above 200 V and a duty cycle (per cent of time electricity is delivered) of >50% (yellow pedal), sparks become electric arcs, and the heat causes the explosion of superficial cells, i.e. cutting. With higher voltages, cutting is associated with coagulation, i.e. blended current. With even higher voltages and a duty cycle <10% preventing arching, only coagulation occurs (blue pedal; forced coagulation). Voltage being crucially important for tissue effects, newer electrosurgical units deliver a constant voltage and limit the energy output (Maximal Watts: W=I*V= joules/sec). Unfortunately, the electrosurgical units indicate the combination of voltage and duty cycles as a force of cutting (pure cutting or blended) or coagulation (soft, forced or spray) current. It is important that the surgeon understands whether electrosurgical units control voltages or output, as well as the electrical basics of the different settings and programs used.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.