Abstract

An electrostatically shielded water-cooled torch was constructed in order to minimize the “secondary discharge” at the sampling orifice, which has been a serious problem in the application of a water-cooled torch to inductively coupled plasma mass spectrometry. The outer gas flow rate could be reduced down to 51/min with an appreciable increase of sensitivity. The intensities of copper ions originating from the sampling orifice decreased to the same level as that of the conventional torch, and doubly charged analyte ions considerably decreased, even compared to those with the conventional torch. A decrease in the plasma potential was also observed with a probe. These facts suggest that the secondary discharge at the sampling orifice was almost completely suppressed by the electrostatic shielding. An increase, however, in the ion signals of analyte oxide was observed. The intensities of ArO+ and Ar2+ decreased, but that of ArN+ increased. The enhancement of ArN+ may have resulted from an entrainment of air into the plasma with the decreasing outer gas flow rate.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.