Abstract
General physicochemical properties of the vascular grafts (VGs) produced from the solutions of Tecoflex (Tec) with gelatin (GL) and bivalirudin (BV) by electrospinning are studied. The electrospun VGs of Tec-GL-BV and expanded polytetrafluoroethylene (e-PTFE) implanted in the abdominal aorta of 36 Wistar rats have been observed over different time intervals up to 24 weeks. A comparison shows that 94.5% of the Tec-GL-BV VGs and only 66.6% of e-PTFE VGs (р = 0.0438) are free of occlusions after a 6 month implantation. At the intermediate observation points, Tec-GL-BV VGs demonstrate severe neovascularization of the VG neoadventitial layer as compared with e-PTFE grafts. A histological examination demonstrates a small thickness of the neointima layer and a low level of calcification in Tec-GL-BV VGs as compared with the control grafts. Thus, polyurethane-based protein-enriched VGs have certain advantages over e-PTFE VGs, suggesting their utility in clinical studies.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.