Abstract
An electrospray ionization (ESI) compatible separation of phospholipids (PL), phosphatidylglycerol (PG), phosphatidylethanolamine (PE), and phosphatidylcholine (PC), was performed on a C18 column by reversed phase High Performance Liquid Chromatography (HPLC) with minimal ESI suppression. The mobile phase, used isocratically, consisted of methanol and water. ESI was used to efficiently transfer the ions present in solution to the gas phase for mass spectrometric (MS) detection. Formation of negative ions was reinforced by incorporating piperidine post column. Limits of detection (LOD) and limits of quantitation (LOQ) were experimentally determined to be 20 and 60 fmol/μl, respectively, when acquiring data in the selected ion monitoring (SIM) mode monitoring three ions with a single quadrupole MS. When acquiring data from m/ z 110–900 in the scanning mode, the LOD and LOQ were experimentally determined to be 1 pmol/μl and 3 pmol/μl. When acquiring product ion spectra for m/ z 747, the LOD and LOQ were experimentally determined to be 446 attomol/μl and 1.3 fmol/μl, respectively.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.