Abstract
A novel and highly sensitive electrospray ionization method, nib spray-mass spectrometry (Nibs-MS), is described. A bamboo pen nib was shaped and used for sample loading and ionization. The nib can be directly connected to the ESI ion sources of various mass spectrometers. The sample solution can be loaded onto the nib by normal pipetting, or simply dipping the nib into a sample solution. The sample solution enters the seam (width ∼7.5μm) of the nib by capillary attraction. Meanwhile, the seam functions as a micro reservoir and prevents the sample solution from spreading to the surface. When a high voltage is applied to the nib, the sample solution is rapidly ejected and ionized toward the mass inlet. Since ionization occurs within a very short period, an abundance of ions is formed, leading to a dramatic improvement in the limit of detection. Using 4-chloro-amphetamine as a model compound, the limit of detection was determined to be 10ng/mL. In contrast to this, when the traditional methods, including paper-spray and wooden tip-spray were used, the limits of detection were 100 and 50ng/mL, respectively. Detailed information on how to prepare a suitable bamboo pen nib, the optimized position for ionization and its application to the analysis of a saliva sample are also reported.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.