Abstract

We describe a method for nanoelectrospray ionization mass spectrometry (nESI-MS) of very small sample volumes. Nanoliter-sized sample droplets were taken up by suction into a nanoelectrospray needle from a silicon microchip prior to ESI. To avoid a rapid evaporation of the small sample volumes, all manipulation steps were performed under a cover of fluorocarbon liquid. Sample volumes down to 1.5 nL were successfully analyzed, and an absolute limit of detection of 105 attomole of insulin (chain B, oxidized) was obtained. The open access to the sample droplets on the silicon chip provides the possibility to add reagents to the sample droplets and perform chemical reactions under an extended period of time. This was demonstrated in an example where we performed a tryptic digestion of cytochrome C in a nanoliter-sized sample volume for 2.5 h, followed by monitoring the outcome of the reaction with nESI-MS. The technology was also utilized for tandem mass spectrometry (MS/MS) sequencing analysis of a 2 nL solution of angiotensin I.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.