Abstract

In electrospray droplet impact (EDI) developed in our laboratory, an atmospheric pressure electrospray source has been used. To increase the ion beam intensity and reduce the evacuation load, a vacuum electrospray cluster ion source using a silica capillary was developed. A silica capillary with a tip inner diameter of 8 µm was used for vacuum electrospray using aqueous 10% methanol. To stabilize the flow rate of the liquid for nano-electrospray, a home-made constant pressure liquid pump was also developed. By using the silica tip nano-electrospray emitter and a constant pressure pump, stable electrospray with flow rate of 22 nL/min was realized without using any heating system such as laser irradiation. Comparative study of mass spectra obtained by atmospheric pressure EDI (A-EDI) and vacuum EDI (V-EDI) was made for various samples such as thermometer molecule, peptide, polystyrene, Alq(3), NPD, C(60), indium, and SiO(2). V-EDI showed slightly milder ionization than A-EDI. Because V-EDI gave higher target current (5-10 nA) than A-EDI (a few nA at most), V-EDI secondary ion mass spectrometry (SIMS) would be a useful technique for the surface and interface analysis.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.