The adsorptive behavior of the immunosuppressive agent tacrolimus was studied by cyclic and differential-pulse voltammetry on a hanging mercury drop electrode (HMDE). The drug was accumulated at HMDE and two well-defined peak currents were obtained at -1353 and -1417 mV vs . SCE (saturated KCl) in borate buffer (pH 10.0) + 0.1 KCl solution. A voltammetric procedure was developed for the determination of tacrolimus using differential-pulse adsorptive stripping voltammetry (DPAdSV). The optimum working conditions for determination of the drug were established. The analysis of tacrolimus in pharmaceutical dosage forms was carried out satisfactorily.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.