Abstract
The pinched injection strategy, implemented on microfabricated fluidic devices (microchips), was investigated for an electrophoretic injection bias. Both the sample loading and dispensing steps were found to contribute to the injection bias whereby neutral species were injected preferentially to anionic species. In the sample loading step, neutral species filled a larger volume in the cross intersection than anionic species. Similarly, in the dispensing step, a larger volume of neutral analyte was injected than anionic analyte. Up to a 27% difference in injected volumes was observed. Fluorescently labeled amino acids were used as model analytes.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.